Title (with LTR): Precision Xceed Glucose Procedure (LTR19678)
Last Approved: Gregory, Kimberly (Electronic Signature Timestamp: 8/11/2014 3:20:23 PM)
4
Comment Codes (Use optional)
COMMENT CODES
0 No Action Required
1 Repeat Test
2 New Test Strip Box
3 New Control Bottle
4 Procedure Error
5 New Test Strip Lot
6 New Control Lot
7 Changed Monitor
8 Changed Batteries
9 Repeat Test, Notify Clinician,
and Send Sample to Lab
11 See Patient Record
15 Training
Test Kits/Supplies/Equipment
Product Manufacturer # Distributed by MGH
PeopleSoft #
Storage
Glucose Precision Xceed
Pro strips
– 100/box
Abbott #70932-01 Owens/Minor 330220 39 - 86°C
Precision Glucose Control
Solution – High/Low, one
bottle of each/box
MediSense #98411-01 Owens/Minor 39961 39 - 86°C
Glucose and Ketone
Calibration verification
Controls (5 levels)
POCT program use only
Abbott #70906-01 N/A
Special request
36 - 86°C
Maintenance
The Precision Xceed Pro™ Monitor requires little routine maintenance. During testing, the sample remains outside the
monitor, which significantly reduces the possibility of contamination.
Disinfecting the Exterior Surface
Disinfecting the exterior surface of the Precision Xceed Pro Monitor is required after each use. The monitor needs only
to be wiped down after use with hospital-approved disinfectant. It is suggested that the monitor be turned off while it is
being cleaned.
DO NOT immerse, flood with any liquid, or autoclave the monitor. Disinfecting after use need not be
documented since this is part of the operating procedure.
Replacing the Batteries (used for replacing alkaline batteries)
Use the following procedure to install new batteries in the Precision Xceed Pro monitor. The Monitor requires two (2)
AA alkaline batteries
1. Turn the Xceed Pro Monitor off by pressing the On/Off button on the keypad.
2. Turn the monitor over to view the battery compartment.
3. Press the tab on the battery compartment cover and lift the cover up.
4. Remove and discard the used batteries.
5. Insert the new batteries, using the + and – symbols in the battery compartment to position the new
batteries with the correct polarity.
Komentáře k této Příručce