Monitor Configuration
The Precision Xceed Pro Monitor configuration settings are enabled via the data management system. These settings
can be viewed on the hand-held device, but cannot be changed on it. Settings that are configured through the data
management system are:
Upload interval
Length of time between data uploads.•
Whether you can continue to test (“allow test” or “warn”) or not (“lockout”).•
Operator ID
Whether the Operator ID is required, optional or not used for all tests.•
Define the Operator ID prompt.•
The minimum and maximum number (between 1 and 30) of digits that are acceptable for the Operator ID. •
With manual entry of the Operator ID, determine whether a check digit is required (modulus 10 or 11).•
Decide acceptable formats for the Operator ID bar code. Refer to the Bar Code Specifications in Chapter 12 for •
available formats.
The monitor can truncate (ignore) the leading digits and/or the last digits of an Operator ID. The monitor can be set •
to select specific digits of an Operator ID. The number of digits is configured in the data management system
.
Patient Tests / Patient ID
Determine whether the Patient ID is required, optional or not used for patient tests.•
Define the Patient ID prompt.•
The minimum and maximum number (between 1 and 30) of digits that are acceptable for the Patient ID.•
Permission for manual entry of the Patient ID.•
Determine whether duplicate entry of a manually entered ID is required.•
With manual entry of the Patient ID, determine whether a check digit is required (modulus 10 or 11).•
Decide acceptable formats for the Patient ID bar code. Refer to the Bar Code Specifications in Chapter 12 for •
available formats.
The monitor can truncate (ignore) the leading digits and/or the last digits of a Patient ID. The monitor can be set to •
select specific digits of a Patient ID. The number of digits is configured in the data management system.
Decide whether confirmation of the Patient ID required.•
Determine how the ID will be confirmed.•
Decide whether comment code is required, optional or not used for patient tests that are in the acceptable range •
for the facility and the test strip.
Decide whether comment code is required, optional or not used for patient tests that are outside of the acceptable •
range for the facility and the test strip.
Control Tests
Set which levels of control solution are required, optional or not used by your facility. The monitor can be •
configured to run any combination of low, mid and high control solution tests.
Set whether control solution test results display as numeric or as Pass/Fail.•
Determine whether a comment code is required, optional or not used for control tests that are within the •
acceptable range for the facility and the test strip.
Decide whether a comment code required, optional or not used for control tests that are outside of the acceptable •
range for the facility and the test strip.
QC Interval
Determine how frequently QC tests need to be run. This can be set to a number of hours or up to three specified •
times during the day.
Decide whether the operator can continue to test (“allow test” or “warn”) or not (“lockout”) when the QC interval •
is expired.